| Toluene | C6H5CH3 | NO.122L | |
|
|
|||

| Performance |
|
Measuring Range |
1 to 2 ppm |
2 to 50 ppm |
50 to 100 ppm |
|
Number of Pump Strokes |
4 |
2 |
1 |
|
Correction Factor |
1/2 |
1 |
2 |
|
Sampling Time |
1.5 minutes per pump stroke |
||
|
Detecting Limit |
0.5 ppm (n=4) |
||
|
Color Change |
White |
||
|
Reaction Principle |
C6H5CH3 + I2O5 + H2SO4 |
||
| Coefficient of Variation | 10% (for 2 to 100 ppm), 5% (for 10 to 50 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Interference |
Change color by itself |
|
Acetylene, Ethylene, Hexane |
No effect |
No discoloration |
|
Alcohols, Ketones, Esters |
No effect |
No discoloration |
|
Aromatics |
Plus error |
Produces brown color |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
Xylene |
4,2,1 |
1,2,4 |
2 to 200ppm |
| Calibration Scale | |
|
Ethyl Benzene Conc. (ppm) |
|
|
Calibration Scale |
|
|
Cumene Concentration (ppm) |
|
| Calibration Scale | |
|
Diethyl Benzene Conc. (ppm) |
|
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 50ppm | - | 1.4 to 6.7% |
|
|