| Trichloroethylene | Cl2C:CHCl | NO.132HA |
|
|
||

| Performance |
|
Measuring Range |
20 to 50 ppm |
50 to 500 ppm |
500 to 1300 ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
0.4 |
1 |
2.6 |
|
Sampling Time |
45 seconds per pump stroke |
||
|
Detecting Limit |
4 ppm (n=2) |
||
|
Color Change |
Yellow |
||
|
Reaction Principle |
Trichloroethylene is decomposed by nascent oxygen by oxidizing agent to |
||
| Coefficient of Variation |
10% (for 50 to 100 ppm), 5% (for 100 to 500 ppm) |
||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Nitric Oxide, Nitrogen dioxide |
- |
No effect |
No discoloration |
|
Hydrogen chloride, Chlorine, Bromine |
- |
Plus error |
Produce reddish purple stain |
|
Acetone |
200 ppm |
No effect |
No discoloration |
|
1,1,1-Trichloroethane |
- |
Plus error |
Produce reddish purple at 3000 ppm |
|
Unsaturated Halogenated HCs |
- |
Plus error |
Produces reddish purple stain |
|
Aromatic hydrocarbons |
100ppm |
Minus error |
No discoloration |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
1,2-Dichloroethylene |
1.6 |
1 |
80 to 800 ppm |
|
1,3-Dichloropropene |
0.9 |
2 |
45 to 450 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 50ppm | 100ppm | - |
|
|