| Trichloroethylene | Cl2C:CHCl | NO.132LL |
|
|
||

| Performance |
|
Measuring Range |
0.125 to 0.25ppm |
0.25 to 4.0ppm |
4.0 to 8.8ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2.2 |
|
Sampling Time |
1.5 minutes per pump stroke |
||
|
Detecting Limit |
0.05 ppm (n=2) |
||
|
Color Change |
Yellow |
||
|
Reaction Principle |
Trichloroethylene is decomposed by nascent oxygen by oxidizing agent |
||
| Coefficient of Variation | 10% (for 0.25 to 1 ppm), 5% (for 1 to 4 ppm) | ||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Hydrogen chloride, Chlorine, |
1/2 time |
Plus error |
Produce purple stain |
|
Tetrachloroethylene |
1/2 time |
Plus error |
Produce purple stain |
|
1,1,1-Trichloroethane |
80 ppm or less |
No effect |
No discoloration up to 80 ppm |
|
Toluene, Xylene |
- |
No effect |
No discoloration |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
1,2-Dichloroethylene |
1.5 |
1 |
0.4 to 6 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 50ppm | 100ppm | - |
|
|