| Vinyl chloride | CH2:CHCl | NO.131L |
|
|
||

| Performance |
|
Measuring Range |
0.1 to 0.2 ppm |
0.2 to 3.0 ppm |
3.0 to 6.6 ppm |
|
Number of Pump Strokes |
4 |
2 |
1 |
|
Correction Factor |
1/2 |
1 |
2.2 |
|
Sampling Time |
1.5 minutes per pump stroke |
||
|
Detecting Limit |
0.02 ppm (n=4) |
||
|
Color Change |
Yellow |
||
|
Reaction Principle |
CH2:CHCl + Cr6+ + H2SO4 |
||
| Coefficient of Variation | 10% (for 0.2 to 1 ppm), 5% (for 1 to 4 ppm) | ||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Ethylene |
200 ppm |
Minus error |
No discoloration |
|
Tetrachloroethylene |
1/3 conc. |
Plus error |
Produce reddish brown |
|
Trichloroethylene |
1/5 conc. |
Plus error |
Produce reddish brown |
|
Benzene, Toluene |
200 ppm |
Minus error |
No discoloration |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
Allyl chloride |
Factor: 16 |
2 |
3.2 to 48 ppm |
|
1,1,2,2-Tetrachloroethane |
Factor: 10 |
2 |
2 to 30 ppm |
| Calibration gas generation | Permeation tube method | |
| TLV - TWA | TLV - STEL | Explosive range |
| 1 ppm | - | 3.6 to 23% |
|
|