| Vinyl chloride | CH2:CHCl | NO.131La | |
|
|
|||

| Performance |
|
Measuring Range |
0.25 to 0.5ppm |
0.5 to 1ppm |
1 to 20ppm |
20 to 54ppm |
|
Number of Pump Strokes |
4 |
2 |
1 |
1/2 |
|
Correction Factor |
1/4 |
1/2 |
1 |
2.7 |
|
Sampling Time |
1 minute per pump stroke |
|||
|
Detecting Limit |
0.05 ppm (n=4) |
|||
|
Color Change |
Yellow |
|||
|
Reaction Principle |
Vinyl chloride is oxidized to form hydrogen chloride to produce red stain. |
|||
| Coefficient of Variation | 10% (for 1 to 6 ppm), 5% (for 6 to 20 ppm) | |||
| Shelf Life | 2Years | |||
| Corrections for temperature & humidity |
Unnecessary | |||
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | ||||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Ethylene |
0.1% |
Minus error |
No discoloration |
|
Tetrachloroethylene |
3 times |
Plus error |
Produce reddish brown |
|
Trichloroethylene |
1/2 time |
Plus error |
Produce reddish brown |
|
Toluene, Xylene |
500 ppm |
Minus error |
No discoloration |
|
Benzene, |
400 ppm |
Minus error |
No discoloration |
| Other substance measurable with this detector tube |
|
Substance |
Correction |
Pump Strokes |
Measuring Range |
|
1,3-Dichloropropane |
Factor: 0.5 |
2 |
0.5 to 10 ppm |
|
p-Ethyl benzylchloride |
Factor: 2.5 |
2 |
2.5 to 50 ppm |
|
Ethyl chloroformate |
Factor: 7 |
2 |
7 to 140 ppm |
|
2-Methyl allyl chloride |
Factor: 2.75 |
1 |
2.8 to 55 ppm |
|
Methyl Chloroformate |
Factor: 58 |
5 |
58 to 1160 ppm |
|
Propylene dichloride |
Factor: 40 |
2 |
40 to 800 ppm |
|
1,2,4-Trichlorobenzene |
Factor: 0.65 |
4 |
0.65 to 13 ppm |
| Calibration gas generation | Permeation tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 1 ppm | - | 3.6 to 23% |
|
|