Water vapor H2O NO.6L


Performance

Measuring Range

0.05 to 1 mg/l

1 to 2 mg/l

Number of Pump Strokes

1

1/2

Correction Factor

1

2

Sampling Time

30 seconds per pump stroke

Detecting Limit

0.03 mg/l (n=1)

Color Change

Yellow Purple

Reaction Principle

Water vapor is absorbed by reagent to produce
alkaline to discolor the indicator to purple.

Coefficient of Variation 10% (for 0.05 to 0.2 mg/l), 5% (for 0.2 to 1 mg/l)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Alcohols

100 ppm

Unclear demarcation

Greenish purple stain
(Whole layer)

Aldehydes

-

Unclear demarcation

Ketones

-

Unclear demarcation

Nitriles

-

Unclear demarcation

Hydrocarbons

-

No effect

No discoloration

Nitrogen dioxide

-

No effect

No discoloration

Sulfur dioxide

100 ppm

Plus error

No discoloration

Ammonia 10 ppm Unclear demarcation Yellowish green


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
- - -


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.