| Xylene | C6H4(CH3)2 | NO.123 | |
|
|
|||

| Performance |
|
Measuring Range |
5 to10 ppm |
10 to 250 ppm |
250 to 625 ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2.5 |
|
Sampling Time |
1.5 minute per pump stroke |
||
|
Detecting Limit |
1 ppm (n=2) |
||
|
Color Change |
White |
||
|
Reaction Principle |
Xylene reacts with iodine pentoxide to |
||
| Coefficient of Variation |
10% (for 10 to 50 ppm), 5% (for 50 to 250 ppm) |
||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Carbon Monoxide |
1000ppm |
2 layers |
Discolors pale brown |
|
Acetylene,Hexane |
2000 ppm |
2 layers |
Discolors pale brown |
|
Toluene |
1/5 time |
Plus error |
Discolors brown |
|
Benzene |
1/5 time |
Plus error |
Discolors pale yellow |
| Other substance measurable with this detector tube |
|
Tube 123 Reading (n = 2) |
10 |
20 |
50 |
100 |
|
Trimethylbenzene Conc. (ppm) |
10 |
27 |
90 |
300 |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 100ppm | 150ppm | 1.0 to 7.0% |
|
|